Technical Specification
Laser Source
Nd:YAG
Wavelength
1064 nanometres
Power to Tissue
1‑12 W in 1 W increment
Pulse Repetition Rate
40 – 60 Hz
Pulse Duration
200 – 400 us.
Fiber
ø 300, 400, or 600 micrometers
Beam Divergence
10 mRAD Minimum, depending of Quality of Fibre Tip
Handpiece
Feed through Handpiece ‑ fully autoclavable
Computerised Operation
Menu selection ‑ treatment settings
Diagnostic Software ‑ Error detection
Cabinet Dimensions
W 35 cm D 40 cm H 90/110 cm
Weight 45 kg
Electrical Requirements
230 Vac / 50 Hz
Consumption max. 7 amps.
Classification
IPX1; Protected against dripping water
Type B equipment
Environment condition
Operation, Transport and Storage condition:
+10oC < t < 40oC
30% < rH < 75%
700hPA < p < 1060 hPA
Approvals
Directive:
MDD 93/42/EEC
Standards:
EN 55 011
EN 60601-1
EN 60825-1
EN 60601-2-22
ISO 13485 – 2003
NB:
G-MED 0459